499: Radiographic Lead Migration in Trial and Permanent Spinal Cord Stimulation
Topic: Clinical Science
Subtopic: Pain category (ICD-11), Other chronic pain category, Neuromodulation (e.g. TMS, TDCS, FUS, SCS, MCS, DBS, vagus nerve stimulation)
Background and Aims: Spinal cord stimulation (SCS) is an established neuromodulation treatment for refractory neuropathic pain conditions. Lead migration can compromise trial interpretation, long-term efficacy, and may lead to revision surgery. Many prior reports describe migration as an incidence-based complication, but this approach may overlook clinically relevant differences in actual displacement magnitude. In particular, trial leads may be vulnerable to early migration because of temporary fixation and externalized hardware, whereas permanent leads may show either excellent stability or rare large-magnitude displacement suggestive of fixation failure. This study compared radiographic maximum migration between trial and conventional permanent SCS leads using millimeter-based measurements and explored whether large-magnitude migration events obscure group-level differences.
Methods: We retrospectively reviewed 148 SCS lead-placement cases at a single institution, including 98 conventional permanent leads and 50 trial leads. Maximum lead migration was measured in millimeters on postoperative radiographic imaging obtained during early follow-up. Because the distribution was non-normal, right-skewed, and included many zero-migration observations, non-parametric statistics were used. Overall trial versus conventional lead migration was compared with the Mann–Whitney U test. A paired subgroup of 26 matched trial–permanent lead pairs from the same patients was analyzed using the Wilcoxon signed-rank test to reduce inter-patient variability. As an exploratory sensitivity analysis, large-magnitude migration was defined as ≥40 mm and excluded to evaluate residual micro-migration patterns.
Results: Among all 148 cases, conventional permanent leads showed a median maximum migration of 0.0 mm and mean migration of 15.26 ± 30.30 mm. Trial leads showed a median migration of 8.9 mm and mean migration of 12.72 ± 17.22 mm. The overall unpaired comparison was not statistically significant (p=0.2814). In the paired subgroup, trial leads migrated significantly more than permanent leads (mean 17.37 mm vs 3.78 mm; p=0.0072). Large-magnitude migration ≥40 mm occurred in 17 cases: 14 conventional leads and 3 trial leads. After excluding these events, the cleaned cohort included 84 conventional and 47 trial leads. In this sensitivity analysis, trial leads demonstrated significantly greater migration than conventional permanent leads (p=0.0143).
Conclusions: Simple unpaired comparison did not detect a significant difference in migration between trial and conventional permanent SCS leads, likely because patient-level variability and large-magnitude displacement events increased dispersion in the overall dataset. However, intra-patient paired analysis demonstrated significantly greater migration during the trial phase, and the same tendency was observed after excluding ≥40 mm events. These findings suggest that millimeter-based radiographic assessment and paired or patient-adjusted analysis better characterize SCS lead stability than incidence-based reporting alone. For patients receiving SCS for refractory neuropathic pain, stricter early trial-phase activity restriction and careful attention to fixation strategy may improve procedural reliability and preserve therapeutic assessment.
References: 1. Shanthanna H, Eldabe S, Provenzano DA, et al. Evidence-based consensus guidelines on patient selection and trial stimulation for spinal cord stimulation therapy for chronic non-cancer pain. Reg Anesth Pain Med. 2023;48(6):273-287. doi:10.1136/rapm-2022-104097.
2. Jenkinson RH, Wendahl A, Zhang Y, Sindt JE. Migration of epidural leads during spinal cord stimulator trials. J Pain Res. 2022;15:2999-3005. doi:10.2147/JPR.S378937.
3. Song WHC, Jen TTH, Osborn JA, Varshney V. Early epidural lead migration in spinal cord stimulator trials: A case series. Interventional Pain Medicine. 2024;3(3):100426. doi:10.1016/j.inpm.2024.100426.
4. Dombovy-Johnson ML, D’Souza RS, Ha CT, Hagedorn JM. Incidence and risk factors for spinal cord stimulator lead migration with or without loss of efficacy: A retrospective review of 91 consecutive thoracic lead implants. Neuromodulation. 2022;25(5):731-737. doi:10.1111/ner.13487.
5. Deer TR, Russo MA, Grider JS, et al. The Neurostimulation Appropriateness Consensus Committee (NACC): Recommendations for surgical technique for spinal cord stimulation. Neuromodulation. 2022;25(1):1-34. doi:10.1016/j.neurom.2021.10.015.
6. Mariadas R, Pan E, Boesel TW, Di Ieva A. Spinal cord stimulators’ displacement: Systematic review and single-institution case series with anchoring technique. J Pain Res. 2026;19:547502. doi:10.2147/JPR.S547502.
7. West T, ElSaban M, Hussain N, et al. Incidence of lead migration with loss of efficacy or paresthesia coverage after spinal cord stimulator implantation: Systematic review and proportional meta-analysis of prospective studies and randomized clinical trials. Neuromodulation. 2023;26(5):917-927. doi:10.1016/j.neurom.2023.03.016.
8. Abraham ME, Potdar A, Ward M, Herschman Y, Mammis A. Risk factors associated with lead migration requiring revision in dorsal root ganglion stimulation. World Neurosurg. 2019;128:e649-e652. doi:10.1016/j.wneu.2019.04.225.
Presenting Author: Jungwon Shin
Submitting Author: Siwook Chung
Senior Author: Siwook Chung
IASP 2026 Global Year theme of Neuropathic Pain: Yes